FDA Clears Expanded Indication for Monitoring Tool in Multiple Myeloma, WM
The decision permits use of the EXTENT system to track monoclonal proteins over time, including from initial diagnosis to ongoing disease management.
The FDA has cleared an expanded intended use for the EXENT Analyser with the EXENT Immunoglobulin Isotypes (GAM) Assay, enabling clinical laboratories to use mass spectrometry–based testing to aid in the monitoring of disease status and treatment response in patients with multiple myeloma and Waldenström macroglobulinemia, according to a press release from the developer, Thermo Fisher Scientific.¹ The expanded clearance builds on the system’s prior 510(k) clearance to aid in the diagnosis of multiple myeloma, which the agency granted in November 2025.2
What did the FDA clear?
The expanded regulatory claim extends the EXENT system, which combines the EXENT Analyser with the GAM assay, from confirmatory diagnosis to ongoing patient monitoring, allowing clinicians to detect and track monoclonal proteins (M-proteins) over time, including at low concentrations. The assay is intended to distinguish disease-related M-proteins from certain known therapeutic antibodies, helping clinicians determine whether a patient with multiple myeloma or Waldenström macroglobulinemia is progressing or responding to treatment.
According to the developers, the clearance makes the EXENT system usable across the full patient journey, from initial diagnosis through longitudinal monitoring, on a single platform. The press release described it as the first mass spectrometry–based monoclonal protein test cleared for ongoing monitoring in the clinical laboratory.
What clinical need does it address, and how does it work?
Multiple myeloma and Waldenström macroglobulinemia are monoclonal gammopathies that require frequent, sensitive monitoring of M-protein levels, typically throughout a patient’s lifetime after diagnosis, to assess disease status, gauge treatment response, and detect potential relapse. Conventional methods such as serum protein electrophoresis (SPEP) and serum immunofixation electrophoresis (sIFE) can be challenged when M-protein concentrations are low or when treatment-related antibodies interfere with interpretation, a growing issue as therapeutic monoclonal antibodies such as daratumumab (Darzalex) and bispecific antibodies are used more widely. The EXENT system uses matrix-assisted laser desorption/ionization time-of-flight (MALDI-TOF) mass spectrometry to identify and quantify M-proteins and is designed for routine laboratory use without prior mass spectrometry expertise, allowing laboratories already using the system for diagnosis to extend the same workflow to monitoring. An estimated 36,000 new cases of multiple myeloma occurred in the US in 2026, per the National Cancer Institute.
“For patients living with multiple myeloma or Waldenström macroglobulinemia, understanding how their disease is responding to treatment is critical,” said Stephen Harding, vice president and general manager of protein diagnostics at Thermo Fisher Scientific, in a news release.1 “This expanded clearance brings the precision of mass spectrometry beyond diagnosis and into ongoing monitoring, giving clinicians greater insight into disease status and treatment response. It reflects our commitment to bringing innovative diagnostics into clinical practice to help physicians make more informed decisions and give patients greater confidence in their care.”
What data support the assay’s performance?
In an extended analytical verification study published in Clinical Chemistry and Laboratory Medicine, the EXENT GAM assay demonstrated acceptable imprecision and a wide linear range (approximately 0.03-30 g/L), with 97% concordance with sIFE for M-protein isotyping and the ability to detect low-level M-proteins below the detection limits of SPEP and sIFE, including some representing residual disease in sIFE-negative samples.3 The assay reliably distinguished endogenous M-proteins from therapeutic monoclonal antibodies, with the exception of talquetamab-tgvs (Talvey), which interfered with quantification and was partially misclassified. The authors also noted that M-protein quantification by the assay differed systematically from SPEP, making the methods non-interchangeable.
The investigators concluded that the assay enables sensitive, reproducible quantification and isotyping of M-proteins below the limits of conventional methods, while adding that further clinical studies are needed to confirm its potential in residual disease detection.
References
- Thermo Fisher Scientific expands clinicians’ ability to monitor multiple myeloma with first-of-its-kind FDA clearance. News release. Thermo Fisher Scientific Inc. November 24, 2026. Accessed November 25, 2026. https://tinyurl.com/45t79vcw
- Thermo Fisher Scientific receives 510(k) clearance in the United States for EXENT system to aid in the diagnosis of multiple myeloma. News release. Thermo Fisher Scientific Inc. November 12, 2025. Accessed November 25, 2026. https://tinyurl.com/3dxttnef
- Nevejan L, Van Droogenbroeck J, Van Hecke E, et al. Extended verification of an automated MALDI-TOF mass spectrometry system for high throughput serum M-protein measurement. Clin Chem Lab Med. 2026;64(4):922-931. doi:10.1515/cclm-2025-0894
Related to this article